A Case in Which the Hole in the Inflation Line Was Not Closed at the Tip of the Tracheal Tube
We report a rare case in which the inflation lumen at the tip of an endotracheal tube (ETT) was open, leading to intraoperative air leakage and cuff deflation. A patient with Down syndrome undergoing planned dental treatment under general anesthesia was induced and nasally intubated with a cuffed ETT that was then inflated with 5 mL of air. Soon thereafter, it was noted that the pilot balloon was deflated and filled with water droplets. The patient was successfully reintubated with a new, replacement ETT. Upon removal, we examined the defective ETT and sent it onward to the manufacturer. Upon further assessment, the manufacturer reported that the inflation lumen was not properly closed during the manufacturing process because of damage that went undetected. Anesthesia providers should assess an ETT for damage prior to use, including ensuring the cuff is functioning properly.

Open Inflation Lumen at the Faulty Endotracheal Tube (ETT) Tip
A. Bubbles emanating from the ETT tip. B. Open end of the inflation lumen at the ETT tip (arrow). C. Enlarged view of the tube tip. The red circle indicated by the black arrow outlines the area of the open inflation lumen.

Components of a Normal or Nonfaulty Endotracheal Tube (ETT)
A-1. Image of the interior of an ETT of the same design showing the inflation lumen (arrow). A-2. Illustration of A-1 showing the inflation lumen (arrow) and the inflation line connection point (blue oval). B-1. Image of the 2 communication holes (arrows) within the ETT cuff. The cuff was broken to enable image capture. B-2. Illustration of B-1. C-1. Image of the ETT tip with the inflation line closed and no observable hole (arrow). C-2. Illustration of C-1. The dotted circle (arrow) outlines the closed end of the inflation lumen at the ETT tip.
Contributor Notes